Integrated poison(special) gas into the system, designed for the cell infection research.
FUNCTION PARAMETER:
• Accurate gas concentration control:
CO2 cell culture mode:
CO2 control range: 0.1% to 20.0%, in 0.1% increments.
Three-gas hypoxic mode:
O2 control range: 0.1% to 23.0%, in 0.1% increments
(can be 0.0% with anaerobic accessory) ;
CO2 control range: 0.1% to 20.0%, in 0.1% increments.
Can set four stages of O2 &CO2 circulation with different concentration.
Cell infection mode:
The air inlet system with quick-connect plugs can inject multiple generators (liquid/solid aerosol generators, VOC generators, smoke generators) into the incubator to carry out cell staining poison test.
• Intimate door-opening gas purge function:
Perform gas purge before opening the door to prevent a large amount of poisonous gas from entering the laboratory.
• Sensor calibration:
One-touch calibration for O2 sensor
• Temperature control:
5 °C above ambient to 45.0° C in 0.1° C increments.
• Humidity control:
Control humidity through the water pan.
• Data Log:
one data set per minute, each set comprises: time, date, O2 (set/actual), CO2 (set/actual).