Descriptions
● one-click start function.
● expandable synchronous external trigger function, which allows external signals (automatic sampler, thermal resolver, etc.) to start the host and workstation simultaneously.
● with a perfect system self-inspection function and automatic fault identification function.
● with 8 external events extension function interface, can be selected with a variety of functional control valves, and work
according to their own set timing.
● 20 groups of samples to test memory function.
● RS232 communication port
Column temperature box:
● content product: 22L
● temperature control range: 5℃ - 400℃ at room temperature
● temperature control accuracy: ±0.1℃
● heating rate: 0.1-60 ℃ / min
● program temperature rise order: 9
● repeatability of programmed temperature rise: ≤ 2%
● cooling way: open the door after
● cooling speed: ≤10 mins (250℃-50℃)
Sampler:
● temperature control range: 7℃ - 420℃ at room temperature
● temperature control mode: independent temperature control
● carrier gas flow control mode: constant pressure
● simultaneous installations: up to 3
● type of injection unit: filling column and triage
● shunt ratio: shunt ratio display
● cylinder pressure range: 0-400kpa
● cylinder pressure control precision: 0.1kpa
● flow setting range: h20-200ml/min and n20-150ml/min
GC112A gas/liquid chromatograph detector:
● choose between FID and TCD
● temperature control range: 7℃ - 420℃ at room temperature
● number of simultaneous installations: up to 2
● ignition function: automatic
● hydrogen flame ionization detector (FID) :
Detection limit: ≤ 3×10-12 g/s (n-hexadecane)
Baseline noise: ≤ 5×10-14 A
Baseline drift: ≤ 6×10-13 A
Dynamic range: 107
RSD: 3% or less
● thermal conductivity detector (TCD) :
Sensitivity: 5000mv. ml/mg (n-hexadecane)
Baseline noise: ≤ 0.05mv
Baseline drift: ≤ 0.15 mV / 30min
Dynamic range: 105
● power supply voltage: 220V±22V, 50Hz± 0.5hz
● power: 3000W