Gas chromatography (GC), is a common type of chromatography used in analytic chemistry for separating and analyzing compounds that can be vaporized without decomposition. Typical uses of GC include testing the purity of a particular substance, or separating the different components of a mixture (the relative amounts of such components can also be determined). In some situations, GC may help in identifying a compound. In preparative chromatography, GC can be used to prepare pure compounds from a mixture
Gas chromatography (GC) is widely used in power sector, corporation that produce oil immersed electrical apparatus--- Transformer , wire cannula and enterprises—power plant, power supply bureau , other units for gas analyzing like chemical processing of coal and Bruckhausen.
Instrument Introduction
GC9790SD gas chromatograph is syclon company for the electric power profession design new generation electrical power system specialpurpose gas chromatograph, uses the most advanced crisis control system and the large screen Chinese display interface, adopt national standard GB/T17623-1998 《In insulating oil dissolved gas component content gas chromatography measuring method》,GB/T7252-2001《In the transformer oil dissolved gas analysis and the judgment guide recommends three detector flows, advanced TCD enlargement technology made to the hydrogen examination ability to enhance 10 times, high sensitive micro current amplifier caused to the hydrocarbon compound gas examination ability to climb to the limit, outstanding chromatographic column system assurance good chromatograph peak shape and the best of high dissociation, stable air circuit control system had guaranteed the analysis accuracy and repeatability, GC9790SD gas chromatograph is the electric power profession user's best choice.
Features:
Advanced TCD enlargement technology, causes to the hydrogen examination ability is common instrumentation's 10 times
High sensitive micro electric current enlargement technology, causes to the hydrocarbon compound gas examination limit ability to climb to the limit
Unique back pressure system design, reduces the examination system's fluctuation enormously, raises the analysis level
Uses three detector flows, two column parallel connection, once injection then realizes 8 kind of gas component automated analysis
Advanced FLASH memory system, causes the data protection no longer to stalemate in battery's expiration
High sensitive PC control system, realizes the online monitoring perfectly
Classical liquid crystal large screen design, parameter setting is simpler intuitively
Perfect self-diagnosis function, breakdown retrieval is no longer complex
Reliable jamproof system, excessively temperature protection function
Technical Specification:
Temperature control range:oven RT+6C-400C,( 1Cincrement)
FID,TCD,Injector,Reburner RT+8C-400C
Accuracy of control temperature:TCD±0.02C Oven,FID,Injector,Reburner±0.1C
FID limit of detection:≤2×10-12g/s[N-C16]
TCD sensitivity:≥8000mV.ml/mg[N-C16]
Minimum detectable concentration:H2≤2ppm CO,CO2≤2ppm C2H2≤0.1ppm
Data processing system:
Power system special purpose Gas Chromatography workstation:WIN9X/XP operating environment, entire Chinese contact surface and operation prompt, automatic computation oil gas content.
Breakdown automatic diagnosis: Provides conforms to transformer oil the dissolved gas analysis guide GB/T7252-2001 three specific value diagnosis reported , prompts the out of limits result automatically, supports TD graphical representation,component density graphical representation etc pattern.
GLP data management: Supports Access database, convenience management,retrieval chromatograph data and other related information.
Application enviroment:
Ambient temperature:5C-35C Relative humidity:less than 85% Power supply voltage:220±10%
Power supply frequency:50±0.5HZ maximum power:2500W